The Combi PAL-xt

The Combi PAL is one of our most popular product to be used with GC systems

Catalogues and Brochures

Detailed Combi PAL-xt brochure


GC accessories catalogue

combi pal catalogue logo  


4 different modes of operation

Liquid mode




combi pal liquid mode
combi pal headspace
combi pal spme mode
combi pal itex


Headspace mode

headspace upgrade

Eliminates dead volume and adsorption effect in sample loops and transfer lines
Permits adjustable sample volumes without sample loop changes
No sample dilution due to vial pressurization

Heated syringe:
+5°C above ambient - 150°C selectable in 1°C increments
Incubator oven:
6 heated vial positions for 2ml/10ml/20ml vials
Incubation temperature:
+5°C above ambient -200°C in 1°C increments



The SPME upgrade


Forensic, food/flavor and pharmaceuticals compounds,characterization of environmental
Variable vial penetration for different types of sample extractions
Sample shaking and heating prior/during extraction – reduction of analysis time

Fiber cleaning:
Optional fiber cleaning station
+5°C above ambient - 350°C, inert gas purging
Incubator oven:
6 heated vial positions for 2ml/10ml/20ml vials
Incubation temperature:
+5°C above ambient -200°C selectable in 1°C increments



The ITEX-2 upgrade

ITEX principle

Get Purge & Trap (P&T) sensitivity without the cost of a P&T System
Rapid & efficient sample enrichment of volatile & semi-volatile compounds in solid, liquid and gaseous
In-tube extraction and direct thermal desorption using proven industry standard adsorbents
Syringe only concept for transparent sample handling, no sample loops, no transfer lines, no switching valves

Extraction Volume: Selectable from 130μI - 1300μI/stroke
Desorption Temperature:
+5°C above ambient - 350°C selectable in 1°C increments
Heating-up rate: up to 12°C/s



MHE (multiple headspace extraction)

The MHE technique requires a series of sampling cycles using the same vial. First, a sample is pressurized to a specific pressure. The vial is then sampled and vented. The vial is re-pressurized and sampled again. This process is repeated multiple times to obtain final results.



Other popular options


Wash stations

wash station wash station large


Flow cell


flow cellflow cell diagram



Solvent dilution

dilutor set combi pal


liquid coolers

liquid coolerdilutor liquid cooler

Tray coolers

tray coolertray cooler 2

Cool stacks


stack coolstack coolstack cool


Heater - agitator

heater agitator




3900 place de Java, Suite 108,  Brossard,  Quebec,  J4Y 0C4,  Canada  (450) 444-0085
Copyright ©
Warning: date(): It is not safe to rely on the system's timezone settings. You are *required* to use the date.timezone setting or the date_default_timezone_set() function. In case you used any of those methods and you are still getting this warning, you most likely misspelled the timezone identifier. We selected the timezone 'UTC' for now, but please set date.timezone to select your timezone. in /services/webpages/l/a/ on line 4